Computerized Lensmeter CL-2800

- Marking & PD Measurement Function
Lens marking and PD measurement functions for rapid final verification - Single Loading Action Printer
A newly designed printer box makes printer paper loading extremely smooth. - Compartment Space
This convenient compartment provides you with extra space to keep printer paper, and the ink cartridge, readily available. - Integrated Transmittance Measuring Function (Factory Option)
To promote multi-coated UV protection on lenses. - The spectral transmittance meter is designed to measure light wavelength transmission in all types of lens material. More importantly, it provides an easy way for the consumer to understand the differences between multi-coated and UV protected sunglasses. A high-tech graphical display helps explain the benefits of UV protection to the eyeglass purchaser. It is ISO compliant covering all wavelengths in standard lens measurements.
Graphic No.1: A single coated normal lens with poor UV-A and B protection.
Graphic No.2: A coated lens with good UV-A and B protection.
| Measuring range | |
|---|---|
| Spherical Power | 0 to ±25D (0.01D/0.12D/0.25D steps) |
| Cylinder Power | 0 to ±10D (0.01D/0.12D/0.25D steps) |
| Cylinder Axis | 0° to 180° (1° step) |
| ADD Power | 0 to +10D (0.01D/0.12D/0.25D steps) |
| Prismatic Power | 0 to 10D(0.12D/0.25D steps) |
| Light Source | LED 620nm |
| Lens Diameter | 5 to 98mm |
| Monitor | 320×240 dot LCD Display Indication: Single/Right-Left/S.C.A. Prism, Addition/Progressive/Abbe. |
| External Interface | RS-232C, USB |
| Power Supply | AC 100, 120, 220 to 240V |
| Power Consumption | 50VA |
| Dimensions | 230(W) × 295(D) × 435(H)mm |
| Weight | 8kg, 10kg (w/transmittance meter) |